Characteristics Under Test
|
Range of Values of Influence Quantities
|
Limits of Variation
|
Pass / Fail
|
|
AC Power
|
102-132 VAC
178-238 VAC
|
Reading cannot vary by more than plus or minus 5%
|
NT
|
|
Battery Power
|
0 - 100 hours
|
Reading cannot vary by more than plus or minus 10%
|
Pass
|
|
Battery Power Indicator
|
Test performed at the voltage that triggers the battery failure indication
|
Reading cannot vary by more than plus or minus 10%
|
NT
|
|
AC powered instrument with battery backup
|
Instrument must be marked for battery endpoint
|
|
Pass
|
|
Test performed at the voltage that triggers the battery failure indication
|
Readings cannot vary by more than plus or minus 10%
|
Pass
|
|
Characteristics Under Test
|
Range of Values of Influence Quantities
|
Limits of Variation
|
Pass / Fail
|
|
Check Circuits
|
Per manufacturer's recommendations
|
|
|
|
Alarms (reset)
|
Dose rate to activate alarm
|
See section 5.2.1
|
Pass
|
|
Alarms (delay)
|
Dose rate to activate alarm
|
Alarm must be indicated within 1 - 60 seconds
|
Pass
|
|
Alarm (threshold drift)
|
Dose rate to activate alarm
|
Alarm setpoint must not drift more than plus or minus 10% over a 500 hour period
|
Pass
|
|
Stability
|
3 hours (battery powered instruments)
|
Reading cannot change by more than plus or minus 6%
|
N/A
|
|
Stability
|
24 hours (AC powered instruments)
|
Reading cannot change by more than plus or minus 6%
|
Pass
|
|
500 hours (AC powered instruments)
|
Reading cannot change by more than plus or minus 15%
|
Pass
|
|
Geotropism
|
Tested in three mutually perpendicular orientations
|
Reading cannot vary by more than plus or minus 6%
|
Pass
|
|
Response Time
|
See Table 1 of Standard
|
See Table 1 of Standard
|
Pass
|
|
Coefficient of Variation
|
Greater than or equal to 1 mR/h, 1 mrd/h, 10 mrem/h, 2000 dpm
|
Reading cannot change by more than plus or minus 10%
|
Pass
|
|
Less than or equal to 1 mR/h, 1 mrd/h, 10 mrem/h, 2000 dpm
|
Reading cannot change by more than plus or minus 15%
|
Pass*
|
|
Line Noise Susceptibility
|
See table 2 of standard
|
Reading cannot change by more than plus or minus 15%
|
NT
|
|
Characteristics Under Test
|
Range of Values of Influence Quantities
|
Limits of Variation
|
Pass / Fail
|
|
Accuracy (photon dose rate)
|
0.1 mrd/h - 1000 rd/h
|
Cannot vary by more than plus or minus 15% of conventionally true value
|
NT
|
|
Accuracy (count rate and contamination monitors)
|
50 dpm/square cm to 100,000 dpm/square cm
|
Cannot vary by more than plus or minus 15% of conventionally true value
|
NT
|
|
Accuracy (beta or neutron dose rate)
|
0.1 mrem/h - 1000 rem/h
|
Cannot vary by more than plus or minus 15% of conventionally true value
|
NT
|
|
Probe surface sensitivity
|
Stated by manufacturer
|
|
NT
|
|
Photon energy dependence
|
80 keV - 1.25 MeV
|
See equation in section 6.3 of standard
|
NT
|
|
20 keV - 3.0 MeV
|
NT
|
|
Beta Energy Dependence
|
0.5 MeV - 3.5 MeV (Emax)
|
See equation in section 6.3 of standard
|
NT
|
|
0.2 MeV - 3.5 MeV (Emax)
|
NT
|
|
Neutron Energy Dependence
|
0.025 eV - 14 MeV
|
See equation in section 6.3 of standard
|
NT
|
|
Photon Radiation Overload
|
100X upper limit less than or equal to 10 rd/h
|
Correct response within 2 minutes
|
NT
|
|
10X upper limit greater than 10 rd/h
|
NT
|
|
Angular Dependence
|
0 - 45 degrees (photon)
|
Instrument reading must not vary by more than plus or minus 20%
|
NT
|
|
45 - 90 degrees
|
Instrument reading must not vary by more than plus or minus 50%
|
NT
|
|
0 - 45 degrees (beta)
|
NT
|
|
Characteristics Under Test
|
Range of Values of Influence Quantities
|
Limits of Variation
|
Pass / Fail
|
|
Temperature
|
0 to 40 degrees C
|
Reading cannot vary by more than plus or minus 15% of reading at 22 degrees C
|
Pass
|
|
-10 to +50 degrees C
|
Reading cannot vary by more than plus or minus 20% of reading at 22 degrees C
|
Pass
|
|
10 to 35 degrees C
|
Reading cannot vary by more than plus or minus 15% of reading at 22 degrees C
|
Pass
|
|
Temperature Shock
|
From -10% to 22 degrees C
|
Pass
|
|
From 50 to 22 degrees C
|
Pass
|
|
Humidity
|
40 to 90% RH at 22 degrees C
|
Readings cannot vary by more than plus or minus 15% of the reading at 40% RH
|
Pass
|
|
Mechanical Shock
|
50 g acceleration of 18 ms, half sine wave, test on 3 orthogonal axes (10 times)
|
Reading cannot vary by more than plus or minus 15%
|
NT
|
|
Vibration
|
2 g acc., 10 - 33 Hz, test on 3 orthogonal axes for 15 min.
|
NT
|
|
Ambient Pressure
|
70 - 106 kPa
|
Pass
|
|
Splashproof
|
2 min. fine spray (4 L/min at 2 meters from nozzle)
|
Pass
|
*Due to the relationship of the response time and the coefficient of variation, readings on the lowest scale
were taken using SLOW response time (manufacturer's suggestion).