Alpha Beta Gamma Detector - Model 43-68
CE Seal


Specifications

INDICATED USE: alpha beta survey
SUGGESTED INSTRUMENTS: Models 12, 16, 18, 2000, 2200, 2221, 2224-1, 2241-2, 2350-1, 2360
DETECTOR TYPE: rechargeable gas proportional
RECOMMENDED COUNTING GAS: P-10 (10% methane, 90% argon)
GAS RECHARGE: will operate on static charge for at least 15 hours with 1 m (39 in.) cable; 5-hour static charge if connected to a simultaneous alpha/beta measuring instrument
WINDOW: 0.8 mg/cm² aluminized Mylar (other thicknesses available)
WINDOW AREA: 126 cm² (19.5 in²) active, 100 cm² (15.6 in²) open
EFFICIENCY (4Ï€): 20%-239Pu; 15%-14C; 30%-99Tc; 30%-90Sr/90Y less than 1%-gamma
SIMULTANEOUS ALPHA/BETA COUNTING: 17.5%-239Pu; 20%-99Tc; 20%90Sr/90Y
BACKGROUND:
alpha: < 3 cpm (when operating at alpha-only plateau region)
beta-gamma: typically 350 cpm (10 μR/hr field)
OPERATING VOLTAGE:
alpha: typically 1100-1400 volts
beta-gamma: typically 1600-1800 volts
COUNTER THRESHOLD SETTING: typically 2-5 mV
GAS CONNECTORS: double-end quick disconnect
SIZE: 10 x 11.7 x 19.8 cm (3.9 x 4.6 x 7.8 in.) (H x W x L), with handle
WEIGHT:0.9 kg (2 lb)

Model 43-68 Probe Gas Accessories
LMI P/N
Quantity
Description
2310017    
1
2-stage Regulator
13-7836
1
Regulator Output Fitting (1/4 in. FPT to ¼ in. tube)
22-9689
3
Plastic Fitting Nuts for ¼ in. vinyl tubing
22-9522
2
Fittings for Flowmeter (1/8 in. MPT to ¼ in. tubing)
21-9547
1
10 to 100 cc Flowmeter (1/8 in. MPT to 1/4 in. tubing)
22-9514
AR
x feet ¼ in. OD Vinyl Tubing for Gas Supply

 

Alpha Beta Gamma Detector