INDICATED USE: Alpha frisking, hand
monitor with alarm
DETECTOR: Alpha air proportional probe with integrated electronics
OPERATION: While sitting in charging/calibration stand, probe may be
used as a hand monitor. Red LED, click audio, and a steady-tone alarm audio
warn of contamination. Probe may also be picked up from charging stand and
used as a frisker for up to 30 minutes before recharge. Maximum recharge
time is 20 minutes to full charge, maximum time to operation is 5 minutes.
DISPLAY: Red LED - Alpha pulse or continuous on alarm
Amber
LED - Trouble
Green LED - OK
WINDOW: 0.4 mg/cm² aluminized mylar with aluminum screen
(79% open).
Can be easily replaced by removal of six screws on back of probe
WINDOW AREA:
Active - 139 cm²
Open - 110 cm²
EFFICIENCY(4pi geometry): Typically 10%- 239Pu
AUDIO: 85 dB at 12 inches
OPERATING VOLTAGE: Altitude sensitive
Sea level - 1950 volts
2000 feet - 1900 volts
5000 feet - 1825 volts
7000 feet -1775 volts
COUNTER THRESHOLD SETTING: 4 mV
CALIBRATION DUE DATE: When exceeded, amber trouble light will illuminate (real
time clock located in charging/calibration stand)
CONSTRUCTION: Conductive plastic body with internal sealed electronics
TEMPERATURE RANGE: -4° F(-20° C) to 122° F(50° C)
May be certified for operation from -40° F(-40° C) to 150° F(65° C)
SIZE: Probe: 9.2” (23.8 cm)H X 3.8” (9.65 cm)W X 1.76” (4.47
cm)D
Charging/Calibration Stand: 0.75” (1.9 cm)H X 0.825” (2.10 cm)W X
1.54” (3.9 cm)L
WEIGHT: 0.75 lbs. (0.34 kg)
POWER: Model 215’s internal capacitors provide power for
30 minutes in
non-alarming condition.
ACCESSORIES:
• Model 215-20 Charging/Calibration Stand interfaces
with a personal computer and uses infrared communications to set parameters and
run automatic
plateau
routines. The stand’s input power is +15V(+12V minimum to +20V maximum)
at 100mA maximum. Four charging/calibration stands can be interconnected to run
from a single 0.5A wall mount power supply.
• Model 215 Calibration Software (Part No.: 1370-080) available to
assist in changing parameters and determining operational settings.
|