PART NUMBER:48-1647 |
||||||||
| INDICATED USE: Gross sample counting COMPATIBLE DETECTORS: G-M, scintillation CONNECTOR: Series "C" (others available) SCALER: 6 digit LED display providing a range of 0 - 999999 counts (controlled by COUNT and HOLD buttons) SCALER LINEARITY: Reading within plus or minus 2% of true value TIMER: Thumb wheel adjustment from 0 - 99 minutes with selectable divisions of X0.1, X1, X10, or EXT for manual timing METER DIAL: 0 - 2.5 kV; BAT TEST LINEARITY: Reading within plus or minus 10% of true value HIGH VOLTAGE: Adjustable from 200 - 2500 volts (will support 60 megohm scintillation loads) THRESHOLD: 50 mV plus or minus 10 mV POWER: 95 - 135 VAC (178 - 240 VAC available), 60 Hz (50 Hz available)single phase (less than 100 mA) METER: 2.5"(6.4cm) arc, 1 mA analog type CONSTRUCTION: Aluminum housing with beige polyurethane enamel paint TEMPERATURE RANGE: -4° F(-20° C) to 122° F(50° C) May be certified for operation from -40° F(-40° C) to 150° F(65° C) SIZE: 8.5"(21.6cm)H X 5"(12.7cm)W X 9.3"(23.5cm)L WEIGHT: 6.5 lbs (3kg) |
|||||||||
|
|
|||||||||
|
|||||||||
|
|
|||||||||
|
Meter Dial |
|||||||||
|
202-014 |
|||||||||
|
|
|||||||||
|
|
|||||||||
|
|
|||||||||
| Ordering Info. | |||||||||
|
|
|||||||||
Copyright © 2000 Ludlum Measurements, Inc. For comments or suggestions please contact webmaster at: ludlum@ludlums.com Page last updated: December 2000 |
|||||||||