INDICATED USE: Field analysis and data logging
COMPATIBLE DETECTORS: G-M, proportional, scintillation
CONNECTOR: Series "C" (others available on request )
DISPLAY: 8 line LCD display with 15 characters per line
BACKLIGHT: A two position toggle switch to turn backlight ON or OFF
SCALER: 6 digit display
TIMER: Used in conjunction with scaler. Count time can be set from 1 - 65,535 seconds in 1 second intervals
RATEMETER: Digital ratemeter, corrected for dead time and calibration constant
TREND INDICATOR: 5 decade logarithmic bar graph
LINEARITY: Reading within plus or minus 10% of true value with detector connected
SCALE: Can display in rem/hr, Sv/hr, R/hr, cpm, cps, dpm, dps, rad(r), Gray(G), C/kg, Ci/cm squared, or
Bq/cm squared
INTEGRATED DOSE: Counter provided to give total accumulated dose for up to 45 days (Will display in same
units as ratemeter )
AUDIO: Built in unimorph speaker with volume control (greater than 60 dB at 2 feet, full volume )
AUDIO DIVIDE: Operator selected divisions of 1, 10, or 100 events-per-click
ALARM: Separate alarms for digital ratemeter, scaler, and integrated dose can be set at any point (audible
and visual indicators )
ACK/SCROLL: Push-button to silence audio after alarm has been indicated and/or scroll through the various
displays
DATA LOGGER: Capable of logging up to 1000 individual data points with the following identifiers for each
point
8 Location codes (5 character ) Time of day
Month, Day, and Year Count rate/scaler count/integrated dose
Count time Detector number
Logging mode Sample number
DETECTOR PARAMETERS: Capable of storing the following parameters for 16 different detectors
Model number Serial number
Calibration constant Dead time correction
High voltage Threshold
Window Display range multiplier
Display time base Display units
Overload current Ratemeter alarm setting
Scaler alarm setting Low ratemeter alarm setting
Integrated dose alarm setting Scaler count time
RS-232 PORT: A full duplex communication port that allows for instrument setup by optional keypad or
PC. Also allows for data to be transferred to a PC file.
BAR CODE READER (optional): Allows for setup of instrument by computer generated bar codes
HIGH VOLTAGE: Adjustable from 200 - 2500 volts
THRESHOLD: Adjustable from 100 - 1000
WINDOW: Adjustable from 0 - 1000 above threshold
GAIN: Adjustable from 2 - 350 mV at threshold setting of 100
DEAD TIME: Adjustable to compensate for dead time of detector and electronics
RESPONSE:
Fixed: - Adjustable from 1 - 127 seconds
in 1 second intervals
Variable: - Will vary according to
number of counts present.
Typical times FAST: 4 - 25 seconds, SLOW: 4 - 60 seconds from 10% to 90% of final reading
POWER: 4 each "D" cell batteries (housed in sealed compartment that is accessible from back
of instrument)
BATTERY LIFE: Greater than 75 hours (low battery condition is automatically indicated )
CONSTRUCTION: Milled and drawn aluminum with beige polyurethane enamel paint
TEMPERATURE RANGE: 32° F(0° C) to 122° F(50° C) (LCD limits temperature range )
SIZE: 6"(15cm)H X 4.3"(11cm)W X 8.8"(22cm)L
WEIGHT: 5 lbs (2.3kg) including batteries
|