PART NUMBER:47-1621


 

The Model 243 is a shielded well counter utilizing a NaI(Tl) scintillator with well.

 
     
     
              
INDICATED USE: Gamma assay
DETECTOR: 1.75"(4.45cm) diameter X 2"(5.08cm) thick integral NaI(Tl) well scintillator
WELL:0.7"(1.7cm) diameter X 1.6"(3.9cm) Deep
EFFICIENCY (4pi geometry): Typically 70% - 129I
SENSITIVITY: Will detect 0.005 microCi gamma or 200 dpm removable contamination
for 99mTc, 131I, 201T, 111In, 125I, 137Cs and 67Ga
COMPATIBLE INSTRUMENTS:
Typically used with Model 1000, Model 2000, Model 2200, and others
TUBE: 2"(5.1cm) diameter magnetically shielded photomultiplier
OPERATING VOLTAGE: Typically 500 - 1200 volts
DYNODE STRING RESISTANCE: 60 megohm
CONNECTOR: Series "C" (others available)
CONSTRUCTION: 0.5"(1.3cm) thick lead and removable cap with beige polyurethane enamel paint
TEMPERATURE RANGE: -4° F(-20° C) to 122° F(50° C)
May be certified for operation from -40° F(-40° C) to 150° F(65° C)
SIZE: 11" (27.9cm)H X 8"(20.3cm)W X 8" (20.3cm)L
WEIGHT: 30 lbs (13.6 kg)
SHIPPING WEIGHT: 46 lbs (20.9kg)

Ordering Info.

Return to ludlums.com