PART NUMBER:48-1410 |
|||||
| INDICATED USE: Beta-gamma survey DETECTORS: Controlled by selector switch INTERNAL: 1" X 1" NaI(Tl) scintillator SENSITIVITY: Typically 175 cpm/microR/hr (137Cs gamma) EXTERNAL: Model 44-38 energy compensated G-M detector (others available) SENSITIVITY: Typically 1200 cpm/mR/hr (137Cs gamma) CONNECTOR: Series "C" (others available) METER DIAL: 0 - 2 mR/hr, 0 - 2.4k cpm, BAT TEST (others available ) MULTIPLIERS: X0.1, X1, X10, X100 LINEARITY: Reading within plus or minus 10% of true value with detector connected AUDIO: Built in unimorph speaker with ON/OFF switch (greater than 60 dB at 2 feet ) CALIBRATION CONTROLS: Accessible from front of instrument (protective cover provided ) HIGH VOLTAGE: Adjustable from 200 - 1500 volts THRESHOLD: 30 mV plus or minus 10 mV RESPONSE: Toggle switch for FAST (4 seconds) or SLOW (22 seconds) from 10% to 90% of final reading RESET: Push-button to zero meter POWER: 2 each "D" cell batteries (housed in sealed compartment that is externally accessible ) BATTERY LIFE: Typically 600 hours with alkaline batteries (battery condition can be checked on meter ) METER: 2.5" (6.4 cm) arc, 1 mA analog type CONSTRUCTION: Cast and drawn aluminum with beige polyurethane enamel paint TEMPERATURE RANGE: -4° F(-20° C) to 122° F(50° C) May be certified for operation from -40° F(-40° C) to 150° F(65° C) SIZE: 9.5"(24.1cm)H X 3.5"(8.9cm)W X 8.5"(21.6cm)L WEIGHT: 4.4 lbs (2 kg) including batteries and detector |
||||||
|
|
||||||
|
Common Meter Dials |
||||||
|
||||||
|
|
||||||
|
|
||||||
|
|
||||||
| Ordering Info.; Manual | ||||||
|
|
||||||
Copyright © 2000 Ludlum Measurements, Inc. For comments or suggestions please contact webmaster at: ludlum@ludlums.com Page last updated: December 2000 |
||||||