PART NUMBER:48-1605

  • 4 Ranges
  • Utilizes G-M, or Scintillation Detectors
  • Typical Counting Range from
    0 - 200 mR/hr or 0 - 500,000 cpm
  • Greater Than 2000 Hour Battery Life

INDICATED USE: General purpose survey
COMPATIBLE DETECTORS: G-M, scintillation
METER DIAL: 0 - 2 mR/hr, or 0 - 5k cpm, BAT TEST (others available)
METER: 2.5" (6.4 cm)arc, 1 mA analog type
MULTIPLIERS: X0.1, X1, X10,X100
RANGE: Typically 0-200 mR/hr, or 0-500,000 counts/minute(cpm)
LINEARITY: Reading within 10% of true value with detector connected
CONNECTOR: Series "C" (others available)
AUDIO: Built in unimorph speaker with ON/OFF switch (greater than 60 dB at 2 feet)
CALIBRATION CONTROLS: Accessible from front of instrument (protective cover provided)
HIGH VOLTAGE: Adjustable from 400 - 1500 volts
THRESHOLD: Fixed at 40 mV ± 10 mV
RESPONSE: Toggle switch for FAST (4 seconds) or SLOW (22 seconds) from 10% to 90% of final reading RESET: Push-button to zero meter
POWER: 2 each "D" cell batteries (housed in sealed compartment that is externally accessible)
BATTERY LIFE: Typically greater than 2000 hours with alkaline batteries (battery condition can be checked on meter)
METER: 2.5" (6.4 cm) arc, 1 mA analog type
CONSTRUCTION: Cast and drawn aluminum with beige powder-coat
TEMPERATURE RANGE: -4°F(-20°C) to 122°F(50°C)
May be certified for operation from -40°F(-40°C) to 150°F(65°C)
SIZE: 6.5" (16.5 cm)H X 3.5" (8.9 cm)W X 8.5" (21.6 cm)L
WEIGHT: 3.5 lbs. (1.6 kg) including batteries

 

 

Suggested detectors are listed below and are not included with the survey meter price. An additional charge will be added for each detector ordered.

 

For
Alpha
Detection

For
Beta
Detection

For
Gamma
Detection

For Alpha/beta/gamma Detection

Model 43-1*
Model 43-2*
Model 43-5*
Model 43-65*
Model 43-90*
Model 44-7
Model 44-9
Model 44-88
Model 44-89
Model 44-94

Model 44-1*
Model 44-6
Model 44-7
Model 44-9
Model 44-21
Model 44-38
Model 44-88
Model 44-89
Model 44-94
Model 44-116*

Model 44-2
Model 44-3
Model 44-6
Model 44-7
Model 44-9
Model 44-10
Model 44-17
Model 44-20
Model 44-21
Model 44-38
Model 133-2
Model 133-4
Model 133-6
Model 133-7
Model 133-8

Model 44-7
Model 44-9
Model 44-88
Model 44-89
Model 44-94

*For Mylar window scintillators it is strongly suggested that the optional overload indicator be used on the Model 3 Survey Meter to better determine the presence of a light leak.

Common Meter Dials

 202-002
0 - 5k cpm
For Any Detector

 202-608
0-6.6k cpm; 0-2 mR/hr
For Model 44-9; 44-9-18;
44-40; 44-88; 44-89; 44-94

 202-664
0-6.6k cpm; 0-20
mSv/h
For Model 44-9; 44-9-18;
44-40; 44-88; 44-89; 44-94


 

 

 

 202-627
0-2 mR/hr

For Model 44-9; 44-9-18;
44-40; 44-88; 44-89; 44-94

202-889
0-20 mSv/h

For Model 44-9; 44-9-18;
44-40; 44-88; 44-89; 44-94
 

202-330
0-4k cpm; 0-2 mR/hr
For Model 44-7
 


 202-676
0-4k cpm; 0-20
mSv/h
For Model 44-7

 202-085
0-2 mR/hr
For Model 44-7

 202-1096
0-20
mSv/hr
For Model 44-7


 

 

202-084
0-2 mR/hr
For Model 44-6; 44-38

202-379
0-20
mSv/h
For Model 44-6; 44-38

202-241
0-2 mR/hr; 0-2.4k cpm
For Model 44-6; 44-38


202-546
0-20
mSv/h; 0-2.4k cpm
For Model 44-6; 44-38

202-654
0-50
mR/hr; 0 - 8.5k cpm
For Model 44-2

202-666
0-50
mR/hr
For Model 44-2


 

 

202-740
0-0.5
mSv/h
For Model 44-2

202-212
0-5
mR/hr; 0-3500 cpm
For Model 44-3(I-125)

202-570
0-50
mSv/h; 0-3.5k cpm
For Model 44-3(I-125)


   

202-717
0-5
mR/hr
For Model 44-10

   

Accessories

Replacement Parts

Check Source
Source Holder
Overload Indicator
Scaler Option
Shoulder Strap
Headset
Standard Carrying Case
Air and Water Tight Carrying Case

Cables
Meter Bezel
Meter Movement
Handle
Probe Clip

Ordering Info.; Technical Info.; Manual

Return to ludlums.com

Copyright © 2000 Ludlum Measurements, Inc.
For comments or suggestions please contact webmaster at: ludlum@ludlums.com