PART NUMBER:48-3458


The Model 375-600 is an area monitor with a 630 in³ shielded plastic scintillation detector .

INDICATED USE:Area monitoring
DETECTOR: 630 in³ shielded plastic scintillation
     DIMENSIONS: 15" (38cm) X 28"(71cm) X 1.5" (3.8cm)
     SHIELDING: 0.125" lead on five sides
SENSITIVITY:30KCPM/µR/hr,137Cs
DISPLAY: 4 digits LED display with 0.8" (2cm) digits
DISPLAY UNITS: kcpm
LINEARITY: Reading within ± 10% of true value with detector connected
RESPONSE: Typically 3 seconds from 10% - 90% of final reading
STATUS: (Green Light) instrument functioning properly
HIGH ALARM: Indicated by red 5 inch diameter xenon strobe, red HIGH ALARM light, and fast beep (4 per second) audible tone (can be set at any point from 0.0 - 9999)
LOW ALARM: Indicated by yellow LOW ALARM light and slow beep (1 per second) audible tone (can be set at any point from 0.0 - 9999)
NOTE: Audible alert and alarm indicators can be configured as a single beep if desired
DET FAIL:(red light and audible tone; greater than 68 dB at 2 feet) indicates detector overload, no count from detector, or instrument failure
LOW BAT: (yellow) indicates less than 2 hours of battery power remaining
DEAD TIME: Adjustable to compenstate for dead time of detector and electronics (can be read on display)
OVERLOAD: Senses detector saturation (indicated by display reading "-OL-")
OVERRANGE: Indicates radiation field being measured has exceeded counting range of instrument (indicated by display reading "____"
DATA OUTPUT: 9 pin connector providing 5 decade logarithmic output, RS-232 output, signal ground connection, FAIL and ALARM signals (current sink), and direct connection to battery and ground, relay outputs available
CALIBRATION CONTROLS: Accessible from front of instrument (protective cover provided)
POWER: 95 - 135 VAC (178 - 240 VAC available), 50-60 Hz single phase (less than 100 mA), 6 volt sealed lead-acid rechargeable battery (built in)
BATTERY LIFE: Typically 8 hours in non-alarm condition
BATTERY CHARGER: Battery is continuously trickle-charged when instrument is connected to line power and turned on
CONSTRUCTION: Steel housing with ivory powder coat paint
TEMPERATURE RANGE: 5° F(-15° C) to 122° F(50° C)
May be certified for operation from -40° F(-40° C) to 150° F(65° C)
SIZE: 2’(61cm)H X 3’(91.5cm)W X 1’(30.5cm)D chassis size
WEIGHT: 125 lbs (57 kg)

Ordering Info.          Manual

Return to ludlums.com

Copyright © 2000 Ludlum Measurements, Inc.
For comments or suggestions please contact webmaster at: ludlum@ludlums.com
Page last updated: May 2005