Part No.: 47-3309

INDICATED USE: Neutron Survey
DETECTOR: PRESCILA Proton Recoil Scintillator
SENSITIVITY: Approximately 350 cpm per mrem/h
NEUTRON ENERGY RESPONSE: Thermal to 100MeV
ANGULAR DEPENDENCE: Within 15% over wide range of energies
GAMMA REJECTION: Approximately 500 cpm at 100mR/h (137Cs)
SUGGESTED INSTRUMENTS: Ludlum Models 12, 2363, 2241 series, and 2350-1
OPERATING VOLTAGE: Typically 500 - 700 volts
CONNECTOR: Series “C” (others available on request)
CONSTRUCTION: Aluminum housing w/black     powdercoat finish
TEMPERATURE RANGE: 14°F (-10°C) to 122°F (50°C)
SIZE: 10.125"(25.7cm)H X 4.25"(10.80 cm)W X 4.25"(10.80 cm)L
WEIGHT: 4.46 lbs.

Ordering Info.        Manual

Return to ludlums.com

Copyright © 2001 Ludlum Measurements, Inc.
For comments or suggestions please contact webmaster at: ludlum@ludlums.com
Page last updated: March 2008