The Model 43-2-2 is a dual phosphor alpha/beta scintillator that is designed to be used for simultaneously counting alpha and beta contamination |
PART NUMBER:47-2003
|
| INDICATED USE: Alpha-beta survey SCINTILLATOR: ZnS(Ag) adhered to 0.010" thick plastic scintillation material WINDOW: Typically 0.8 mg/cm² aluminized mylar (1.2 mg/cm² recommended for outdoor use) WINDOW AREA: Active and Open - 12 cm² EFFICIENCY (4pi geometry): Typically 30% - 239Pu; 30% - 230Th; 20% - 90Sr/90Y; 5% - 14C BACKGROUND: Alpha - Less than 3 cpm Beta - Typically 50 cpm or less (10 microR/hr field) NON-UNIFORMITY: Less than 10% CROSS TALK: Alpha to Beta - Less than 10% Beta to Alpha - Less than 1% COMPATIBLE INSTRUMENTS: Model 2224, Model 2360, Model 2929 TUBE: 1.5"(3.8cm) diameter magnetically shielded photomultiplier OPERATING VOLTAGE: Typically 500 - 1200 volts DYNODE STRING RESISTANCE: 100 megohm CONNECTOR: Series "C" (others available) CONSTRUCTION: Aluminum housing with beige polyurethane enamel paint TEMPERATURE RANGE: -4° F(-20° C) to 122° F(50° C) May be certified for operation from -40° F(-40° C) to 150° F(65° C) SIZE: 2"(5.1cm) diameter X 7.3"(18.5cm)L WEIGHT: 1 lb (0.5kg) |
|
|
|
|
|
Replacement Parts |
|
|
|
|
| Ordering Info. Manual | |
|
|
|
Copyright © 2000 Ludlum Measurements, Inc. For comments or suggestions please contact webmaster at: ludlum@ludlums.com Page last updated: December 2000 |
|