The Model 43-44 is a 50 cm² air proportional alpha detector for alpha survey.

PART NUMBER:47-1169

INDICATED USE: Alpha survey
WORKING ENVIRONMENT: Splashproof shields and desiccant vented chamber for outdoor use
WINDOW: 0.4 mg/cm² aluminized mylar
WINDOW AREA:
     Active - 76 cm²
     Open - 50 cm²
EFFICIENCY(4pi geometry): Typically 8%- 239Pu
COMPATIBLE INSTRUMENTS: Models 12, 16, 18, 2000,2200, 2221, 2241, 2350
OPERATING VOLTAGE: Altitude sensitive
     Sea level - 2050 volts
     2000 feet - 2000 volts
     5000 feet - 1925 volts
     7000 feet - 1875 volts
COUNTER THRESHOLD SETTING: 2 mV
CONNECTOR: Series "C"(others available )
CONSTRUCTION: Anodized aluminum body with aluminum screen (65% open)
TEMPERATURE RANGE: -4° F(-20° C) to 122° F(50° C)
May be certified for operation from -40° F(-40° C) to 150° F(65° C)
SIZE: 4" (10.2 cm)H X 2.5" (6.4 cm)W X 9" (22.9 cm)L
WEIGHT: 1 lb (0.5kg)

Replacement Parts

Mylar Window
Electrodes
Dehydrators

Ordering Info.        Manual

Return to ludlums.com

Copyright © 2000 Ludlum Measurements, Inc.
For comments or suggestions please contact webmaster at: ludlum@ludlums.com