The Model 44-6 is a sidewall G-M detector with a rotary beta shield for general purpose survey.

PART NUMBER:47-1535

INDICATED USE: Beta gamma survey
DETECTOR: 30 mg/cm² stainless steel wall halogen quenched G-M
SENSITIVITY: Typically 1200 cpm/mR/hr (137Cs gamma )
BETA CUT OFF: Approximately 200 keV (window open )
ENERGY RESPONSE: Energy dependent
DEAD TIME: Typically 95 microseconds
COMPATIBLE INSTRUMENTS: General purpose survey meters, ratemeters, and scalers
OPERATING VOLTAGE: 900 volts
CONNECTOR: Series "C" (others available )
CONSTRUCTION: 1000 mg/cm² polished stainless steel housing with rotary beta window
TEMPERATURE RANGE: -4° F(-20° C) to 122° F(50° C)
May be certified for operation from -40° F(-40° C) to 150° F(65° C)
SIZE: 1.2" (3.0 cm) diameter X 6.5" (16.5 cm)L
WEIGHT: 1 lb (0.5kg)

Replacement Parts

G-M Tube

 

Response Curves

                       

        Model 44-6 Closed Response Curve                    Model 44-6 Open Response Curve

 Ordering Info.        Manual

Return to ludlums.com