The Model 44-88 is a cylindrical pancake G-M detector for contamination survey. |
PART NUMBER:47-2356 |
| INDICATED USE: Alpha beta gamma survey; Sample counting DETECTOR: Pancake type halogen quenched G-M WINDOW: 1.7 ± 0.3 mg/cm² mica WINDOW AREA: Active - 15 cm² Open - 12 cm² EFFICIENCY(4pi geometry): Typically 5%-14C; 22%-90Sr/90Y; 19%-99Tc; 32%-32P; 15%-239Pu SENSITIVITY: Typically 3300 cpm/mR/hr (137Cs gamma ) ENERGY RESPONSE: Energy dependent DEAD TIME: Typically 80 microseconds COMPATIBLE INSTRUMENTS: General purpose survey meters, ratemeters, and scalers OPERATING VOLTAGE: 900 volts CONNECTOR: Series "C" (others available ) CONSTRUCTION: Aluminum housing with beige polyurethane enamel paint TEMPERATURE RANGE: -4° F(-20° C) to 122° F(50° C) May be certified for operation from -40° F(-40° C) to 150° F(65° C) SIZE: 2.3" (5.7 cm) Diameter X 2.8" (7 cm)L WEIGHT: 0.5 lb (0.2kg) |
|
|
|
|
|
Replacement Parts |
|
|
|
|
| Ordering Info. | |
|
|
|