The Model 44-88 is a cylindrical pancake G-M detector for contamination survey.

PART NUMBER:47-2356

INDICATED USE: Alpha beta gamma survey; Sample counting
DETECTOR: Pancake type halogen quenched G-M
WINDOW: 1.7 ± 0.3 mg/cm² mica
WINDOW AREA:
     Active - 15 cm²
     Open - 12 cm²
EFFICIENCY(4pi geometry): Typically 5%-14C; 22%-90Sr/90Y; 19%-99Tc; 32%-32P; 15%-239Pu
SENSITIVITY: Typically 3300 cpm/mR/hr (137Cs gamma )
ENERGY RESPONSE: Energy dependent
DEAD TIME: Typically 80 microseconds
COMPATIBLE INSTRUMENTS: General purpose survey meters, ratemeters, and scalers
OPERATING VOLTAGE: 900 volts
CONNECTOR: Series "C" (others available )
CONSTRUCTION: Aluminum housing with beige polyurethane enamel paint
TEMPERATURE RANGE: -4° F(-20° C) to 122° F(50° C)
May be certified for operation from -40° F(-40° C) to 150° F(65° C)
SIZE: 2.3" (5.7 cm) Diameter X 2.8" (7 cm)L
WEIGHT: 0.5 lb (0.2kg)