PART NUMBER:48-1607 |
||||||
|
||||||
| INDICATED USE: Gamma survey DETECTORS: Two internal energy compensated G-M tubes ENERGY RESPONSE: Within plus or minus 15% of true value from 60 keV - 3 MeV SATURATION: In excess of 1000 R/hr METER DIAL: 0 - 2 mR/hr, BAT TEST (others available) MULTIPLIERS: X0.1, X1, X10, X100, X1000 LINEARITY: Reading within plus or minus 10% of true value with detector connected AUDIO: Built in unimorph speaker with ON/OFF switch (greater than 60 dB at 2 feet ) CALIBRATION CONTROLS: Accessible from front of instrument (protective cover provided ) HIGH VOLTAGE: Adjustable from 200 - 1500 volts THRESHOLD: 30 mV plus or minus 10 mV RESPONSE: Toggle switch for FAST (4 seconds) or SLOW (22 seconds) from 10% to 90% of final reading RESET: Push-button to zero meter POWER: 2 each "D" cell batteries (housed in sealed compartment that is externally accessible ) BATTERY LIFE: Typically 600 hours with alkaline batteries (battery condition can be checked on meter ) METER: 2.5" (6.4 cm) arc, 1 mA analog type CONSTRUCTION: Cast and drawn aluminum with beige polyurethane enamel paint TEMPERATURE RANGE: -4° F(-20° C) to 122° F(50° C) May be certified for operation from -40° F(-40° C) to 150° F(65° C) SIZE: 9.5"(24.1cm)H X 3.5"(8.9cm)W X 8.5"(21.6cm)L WEIGHT: 4.4 lbs (2 kg) including batteries and detector |
||||||
|
|
||||||
|
Common Meter Dial |
||||||
|
|
||||||
|
|
||||||
|
|
||||||
|
|
||||||
| Ordering Info. Manual Model 5 Response Curve | ||||||
|
|
||||||
Copyright © 2000 Ludlum Measurements, Inc. For comments or suggestions please contact webmaster at: ludlum@ludlums.com Page last updated: December 2000 |
||||||