PART NUMBER:48-1608

  • Ion Chamber
  • 4 Ranges
  • Retractable Beta Shield
  • Total Counting Range from
    0 - 5000 mR/hr
  • Built-In Audio
 
INDICATED USE: General purpose survey
CHAMBER CONSTRUCTION: Carbon coated acrylic
CHAMBER VOLUME: 220 cm³
WINDOW: 7 mg/cm² aluminized mylar
WINDOW AREA: Active - 40 cm²
                                Open - 30 cm²
BETA SHIELD: Retractable 300 mg/cm² phenolic slide
ENERGY RESPONSE: Within plus or minus 20% of true value from 10 keV - 2 MeV
BETA FACTOR: 5.4 with uranium slab
AUDIO: Built in unimorph speaker with ON/OFF switch
METER DIAL: 0 - 5 mR/hr, BAT TEST
MULTIPLIERS: X1, X10, X100, X1000
LINEARITY: Reading within plus or minus 10% of true value
CALIBRATION CONTROLS: Accessible from front of instrument (protective cover provided)
RESPONSE: Typically 3 seconds from 10% to 90% of final reading
RESET: Push-button to discharge chamber and zero meter
ZERO ADJUST: 3 turn potentiometer
WARM UP TIME: Approximately 5 minutes
POWER: 2 each "D" cell batteries (housed in sealed compartment that is accessible from front of instrument)
BATTERY LIFE: Greater than 200 hours with alkaline batteries (battery condition can be checked on meter)
METER: 2.5" (6.4 cm) arc, 1 mA analog type
CONSTRUCTION: Cast and drawn aluminum with beige polyurethane enamel paint
TEMPERATURE RANGE: -4° F(-20° C) to 122° F(50° C)
May be certified for operation from -40° F(-40° C) to 150° F(65° C)
SIZE: 8"(20.3 cm)H X 3.5"(8.9 cm)W X 8.5"(21.6 cm)L
WEIGHT: 4 lbs (1.8 kg) including batteries

Common Meter Dial

202-200
0 - 5 mR/hr

202-252
0 - 50
mSv/h

Accessories

Replacement Parts

Check Source
Headset
Standard Carrying Case
Air and Water Tight Carrying Case

Meter Bezel
Meter Movement
Handle
Ion Chamber
Dehydrators

Ordering Info.          Manual

Return to ludlums.com

Copyright © 2000 Ludlum Measurements, Inc.
For comments or suggestions please contact webmaster at: ludlum@ludlums.com